Prosthetic valve evaluation with low-dose dobutamine

Stress echoAdvanced
Flowchart: Prosthetic valve evaluation with low-dose dobutamine. A text version follows the figure.
Text of this figure
  • Prosthetic valve evaluation with low-dose dobutamine
  • ΔQ ≥20%
  • ΔQ <20%
  • Stress EOA
  • ≪ normal EOA
  • Stress EOA
  • ~ normal EOA
  • Small prosthetic valve EOA (<1.0 cm²) and/or abnormal DVI (<0.35 aortic; >2.2 mitral)
  • Low flow (SVi <35 mL/m²) and symptoms
  • Low-dose dobutamine-stress echo
  • LV flow reserve
  • No LV flow reserve
  • ΔMPG >20 mm Hg (aortic)
  • ΔMPG >10 mm Hg (mitral)
  • ΔEOA <0.3 cm²
  • ΔMPG <20 (aortic)
  • ΔMPG <10 (mitral)
  • ΔEOA ≥0.3 cm²
  • Indeterminate
  • True significant
  • dysfunction or PPM
  • Pseudo dysfunction
  • or PPM
  • Dysfunction ± PPM
  • Isolated PPM
  • Source: The Clinical Use of Stress Echocardiography in Non-Ischaemic Heart Disease: Recommendations from the European Association of Cardiovascular Imaging and the
  • American Society of Echocardiography (2017), Figure 23.