Prosthetic mitral valve regurgitation

Prosthetic and repaired valvesAdvanced
Flowchart: Prosthetic mitral valve regurgitation. A text version follows the figure.
Text of this figure
  • Prosthetic mitral valve regurgitation
  • Yes, mild*
  • Yes, severe*
  • **
  • 2–3
  • criteria
  • 2–3
  • criteria
  • ≥4 criteria
  • Definitely mild
  • ≥4 criteria
  • Definitely severe
  • 3 specific criteria
  • for severe MR
  • Prosthetic mitral valve regurgitation
  • Does MR meet specific criteria for
  • mild or severe MR?
  • Specific criteria for mild MR
  • • Small, narrow central jet
  • • VCW ≤0.3 cm
  • • VCA <0.2 cm²
  • • PISA radius absent or ≤0.3 cm at Nyquist
  • 30–40 cm/s
  • • Mitral A-wave dominant inflow
  • • Soft or incomplete jet by CW Doppler
  • Specific criteria for severe MR
  • • Valve dehiscence, incomplete occluder
  • closure, bioprosthetic flail leaflet
  • • VCW ≥0.7 cm or ≥2 moderate jets
  • • VCA (1 or 2 jets) ≥0.4 cm²
  • • PISA radius ≥1.0 cm at Nyquist 30–40
  • cm/s**
  • • Central large jet >50% of LA area
  • • Pulmonary vein systolic flow reversal
  • Intermediate values:
  • MR probably moderate
  • Perform quantitative methods whenever possible
  • to refine assessment
  • RVol <30 mL
  • RF <30%
  • **EROA <0.2 cm²
  • RVol 30–59 mL¶
  • RF 30–49%
  • **EROA 0.2–0.39 cm²
  • RVol ≥60 mL¶
  • RF ≥50%
  • **EROA ≥0.4 cm²
  • Mild MR
  • Moderate MR
  • Severe MR
  • • Poor TTE quality or low confidence in measured Doppler parameters
  • • Discordant quantitative and qualitative parameters and/or clinical data
  • Indeterminate MR
  • Consider further testing:
  • TEE or CMR for quantitation
  • * Beware of underestimation of MR severity by color Doppler on TTE because of flow masking, particularly in mechanical valves.
  • Beware of underestimation of severity by color Doppler in eccentric, wall-impinging jets.
  • ** EROA by PISA may overestimate severity if PISA is ovoid shape, constrained by the LV wall.
  • ¶ Regurgitant volume cutoffs for severe MR may be lower in low-flow conditions.