Spectral Doppler artifacts: mirroring, double envelope, tiger stripes

ViewPW and CW recordings, especially near the pleura or pericardium, across prosthetic or calcified valves, and over oscillating structures.
TimingMirrored signals are simultaneous with the true signal; tiger stripes are horizontal bands on both sides of the baseline.
Caliper / traceMeasure only the dense true envelope: the mirrored copy is less dense on the opposite side; the double envelope is a faint outer copy at exactly twice the modal velocity.
No image yet
No cleared image or clip is available for this item yet; the description below stands in. Fabricated or stock images are not used.

Correct example

A faint symmetric signal below the baseline of pulmonary venous or mitral flow is a spectral mirror; a faint outer envelope at twice the velocity of the dense CW jet is a double envelope; evenly spaced bands across the baseline are tiger stripes from an oscillating leaflet, lead or mass.

Common error

Tracing the outer duplicated envelope (doubling the gradient), reading a mirror signal as reversed flow, or interpreting tiger stripes as a flow signal.

Effect on interpretation

A doubled envelope overestimates velocity and gradient across a prosthetic or calcified valve; mirroring suggests non-existent bidirectional flow. Lowering Doppler gain and improving alignment reduce both.

Why

Mirroring arises from near-perpendicular interrogation or system misinterpretation; the double envelope is a reverberation/duplication from a strong reflector in high flow with high gain; tiger stripes are the fundamental and harmonic frequencies of a vibrating structure.