LVOT diameter and sample position

ViewZoomed parasternal long axis showing the centre axis of the LVOT and the aortic cusp insertion points.
TimingMid-systole.
Caliper / traceInner edge to inner edge, compacted septum to anterior mitral leaflet, about 3–10 mm from the valve plane, at the same position as the PW sample volume used for the LVOT VTI.
Example
Schematic parasternal long-axis outline with a caliper across the LVOT 3–10 mm below the aortic valve plane.
Caliper position in the LVOT relative to the valve plane. (Schematic drawing (E/e′ureka), not a patient image)

Correct example

Zoomed PLAX, gain adjusted to the blood–tissue interface, caliper placed perpendicular to the outflow tract 3–10 mm below the cusp insertions in mid-systole; the PW sample volume for the VTI is then placed at that same distance from the valve.

Common error

Measuring the aortic annulus (cusp insertion) or an oblique, non-zoomed image and pairing it with a VTI sampled at a different level; measuring a single beat in an irregular rhythm.

Effect on interpretation

The diameter is squared to obtain a circular area, so a small error in diameter is magnified in stroke volume and continuity AVA; underestimating the LVOT area is the most important source of AVA underestimation.

Why

Stroke volume is CSA × VTI, and CSA assumes a circle from the diameter; the diameter and the velocity sample must come from the same level of the outflow tract for the product to describe one flow.